Logo specs-groupDetail

HIS 14 HD

The HIS 14 is an advanced, yet easy to use small spot UV light source with long time proven design. 

The HIS 14 HD lamp is based on the HIS 13 with an additional focussing mirror. The focussing mirror provides ideal conditions for high resolution ARPES: a spot size of 180 µm enabling high angular resolution with classical ARPES analyzers and at the same time delivering a photon density which allows for angular resolved photoemission spectroscopy close to synchrotron performance level. 

KEY FEATURES

  • Long lifetime due to filament free design
  • 180 µm spot size
  • photon density up to  1.1·1013 Ph/mm2/s
  • Large working distance (up to 200 mm)​
  • line of sight viewport
  • Operating pressure in the 10-10 mbar range
  • Ease of operation

The monochromator delivers pure He I – w/o satellites or He II – lines for ghost peak free angle resolved photo electron spectoscopy (ARPES) or momentum microscopy. It is an option for the HIS 13 as well as the HIS 14 HD. For more details please refer to the dedicated section for HIS Mono.

The HIS 14 HD can be focused down to 180 µm or better for a high photon flux density and small emission spot. Conventional analysers benefit from a small photo-electron emission spot equally as microscopes with an intrinsic small field of view. 

Using a conventional ARPES analyser, a small spot UV source increases the effective angular resolution, providing sharper band strutures. Small spot analysers and microscopes benefit from the increased photo-electron yield, if the light spot is equal or smaller than the field of view. In this case no artificial loss due to undetected photo-electrons is observed. 

Easy Integration

The HIS 14 HD, can be easily retrofitted in various chambers and designs. The long focusing length and the source geometry ensure an easy integration into most existing analyzer setups. 

MADE FOR THESE METHODS

3

SPECIFICATIONS

HIS 14 HD
Operation
Operating Gases

He, Ne, Ar, Kr, Xe, H

Working Pressure

< 5x10-8 mbar (2 Stage)
< 1x10-9 mbar (3 stage)

Working Distance

70 mm

Working Principle

High Voltage Discharge

Performance
Photon Flux Density
  • 4x1012 ph/(s*mm²)
  • 8x1011 ph/(s*mm²) with HIS Mono
Spot Size

<180mm

Capillary

Focussing Mirror

Mounting
Mounting Flange

DN63CF (DN100CF)