Logo specsDetail

PHOIBOS 150 NAP 2D-CMOS

State of the art hemispherical energy analyzer with 2D-CMOS detector for photoelectron spectroscopy measurements (XPS and UPS) in the pressure regime from UHV to near ambient pressure (NAP)

The PHOIBOS 150 NAP Analyzer is a true 180° hemispherical energy analyzer with 150 mm mean radius. It consists of a differentially pumped electrostatic pre-lens, with a three-stage differentially pumped PHOIBOS 150 analyzer. Thus, the design concept provides four separate pressure stages separated by apertures. The first pumping stage (pre-lens) is separated from the analytic chamber by a nozzle with a customizable opening at the tip with diameters between 0.3 mm and 1 mm. By using a turbopump on the pre-lens stage, a pressure difference of four orders of magnitude (compared to the analysis chamber) can be achieved.The first and second stages are separated by an aperture. An in-lens gate valve allows a high-vacuum seal between of the PHOIBOS 150 Analyzer and NAP pre-lens and enables venting of the analysis chamber and pre-lens without venting the energy analyzer.

For Imaging NAP-XPS the pre-lens can be equipped with the SPECS NAP-XPS Imaging Lens Module that supports two different operation modes, one optimized for lateral resolution and one for transmission. In the lateral resolving mode the acceptance angle can be freely adjusted between +/- 3° to +/- 8° giving an ultimate lateral resolution of better than 10 µm with an acceptance area of 0.6 mm in diameter.

KEY FEATURES

  • Wide Angle Pre-Lens with 44 ° Acceptance Angle
  • Near Ambient Working Pressures up to 100 mbar (depending on configuration)
  • Large pass energy range
  • Working range up to 3.5 keV (upgradable to 7 keV with corresponding power supply and detector)
  • High energy and angular resolution
  • High spatial resolution with imaging lens module
  • Ultimate flexibility by switching between high transmission and high lateral resolution mode
  • Fast detectors with high dynamic range for fast real time data acquisition in snapshot mode
  • Pneumatic in-lens gate valve
  • 4 differential pumping stages
  • Full detector flexibility (MCD, 1D-DLD, 2D-DLD, 2D-CMOS)

R&D100 in 2010
The PHOIBOS 150 NAP Analyzer won the R&D 100 award for the best 100 products developed in 2010.

MADE FOR THESE METHODS

1

SPECIFICATIONS

PHOIBOS 150 NAP 2D-CMOS
Operation
Energy Dispersion

Hemisphere

Slits/Apertures

8 Entrance, 3 Exit slits and Iris aperture

Magnetic Shielding

Double µ-Metal Shielding

Lens Modes

Angular Resolving, Transmission and Magnification (Lateral Resolution) Lens Modes

Kinetic Energy Range

0 - 3500 eV

Pass Energies

0 - 550 eV continously adjustable

Detector

2D-CCD/CMOS Detector

Measurement Modes

Snapshot mode, Sweeping mode, Fixed energy mode

Energy Window

13% of Pass Energy

Electric Isolation

up to 7 keV

Electronics

HSA 3500 plus HT 101 for analyzer and HSA 3500 plus HT 173 for pre-lens

Working Pressure

up to 30 mbar (higher pressure values achievable with corresponding nozzle diameters and differential pumping packages)

Performance
Acceptance Angle

±22°

Angular Resolution

< 0.7°

Energy Resolution

< 10 meV

Lateral Resolution

<15 μm (guaranteed), <10 μm (achievable) with imaging lens module

Smallest Acceptance Spot

Without imaging lens module the smallest acceptance spot is defined by the excitation spot of X-ray source or nozzle diameter

Detector Channels

1900 x 1200 (with channel binning)

XPS Count Rates UHV

70 kcps (guaranteed), 150 kcps (achievable) *

XPS Count Rates 10 mbar

7 kcps (guaranteed), 15 kcps (achievable) *

XPS Count Rates 25 mbar

0.5 kcps (guaranteed), 1.5 kcps (achievable) *

.

* Ag 3d, FWHM < 0.85 eV, small spot monochromated X-ray source µ-FOCUS 600, Al Kα anode, 20 W, Spot size < 250 µm

Mounting
Mounting Flange

DN150 CF (8" OD)

Working Distance

300 - 500 µm (for standard nozzle with 300 µm diameter)

RELATED PRODUCTS

18

APPLICATION NOTES

PUBLICATIONS

  1. (2021) A comparative study of electrochemical cells for in situ x-ray spectroscopies in the soft and tender x-ray range

    n situ x-ray spectroscopies offer a powerful way to understand the electronic structure of the electrode–electrolyte interface under operating conditions. However, most x-ray techniques require vacuum, making it necessary to design spectro-electrochemical cells with a delicate interface to the wet electrochemical environment. The design of the cell often dictates what measurements can be done and which electrochemical processes can be studied. Hence, it is important to pick the right spectro-electrochemical cell for the process of interest. To facilitate this choice, and to highlight the challenges in cell design, we critically review four recent, successful cell designs. Using several case studies, we investigate the opportunities and limitations that arise in practical experiments.



    J.-J. Velasco-Vélez, L. J. Falling, D. Bernsmeier, M. J Sear, P. C. J. Clark, T.-S. Chan, E. Stotz, M. Hävecker, R. Kraehnert, and A. Knop-Gericke
    Juan-Jesús Velasco-Vélez et al 2021 J. Phys. D: Appl. Phys. 54 124003
    Read more
  2. (2021) In situ investigation of the bismuth vanadate/potassium phosphate interface reveals morphological and composition dependent light-induced surface reactions

    Bismuth vanadate (BiVO4) is an established n-type oxide semiconductor for photoelectrochemical oxygen evolution. Direct charge carrier recombination at the solid/liquid interface is a major cause of efficiency loss in BiVO4-based devices. Intrinsic and extrinsic surface states (SSs) can act as electron and hole traps that enhance the recombination rate and lower the faradaic efficiency. In this study, we investigate the BiVO4/aqueous KPi interface using two types of samples. The samples were prepared at two different deposition and annealing temperatures (450 °C and 500 °C) leading to different morphologies and stoichiometries for the two samples. Both samples exhibit SSs in the dark that are passivated under illumination. In situ ambient pressure hard x-ray photoelectron spectroscopy experiments performed under front illumination conditions reveal the formation of a bismuth phosphate (BiPO4) surface layer for the sample annealed at 450 °C, whereas the sample annealed at 500 °C exhibits band flattening without the formation of BiPO4. These results imply that the light-induced formation of BiPO4 may not be responsible for SS passivation. Our study also suggests that slight differences in the synthesis parameters lead to significant changes in the surface stoichiometry and morphology, with drastic effects on the physical-chemical properties of the BiVO4/electrolyte interface. These differences may have important consequences for device characteristics such as long-term stability.



    M. Favaro, I.Y. Ahmet, P. C. J. Clark, F. F. Abdi, M. J. Sear, R. van de Krol, and D. E. Starr
    Marco Favaro et al 2021 J. Phys. D: Appl. Phys. 54 164001
    Read more
  3. (2019) New Insight into the Gas-Sensing Properties of CuOx Nanowires by Near-Ambient Pressure XPS

    This article presents an investigation of the sensing properties of chemiresistors based on Cu2O/CuO core–shell nanowires containing p–p′ heterojunctions. The nanowires were synthesized by a conventional hydrothermal method and used for the detection of ethanol and nitrogen dioxide, reducing and oxidizing agents, respectively. To unravel the chemical processes connected with gas detection, an in situ approach was applied. This approach was based on near-ambient pressure X-ray photoelectron spectroscopy combined with simultaneous monitoring of sensor responses. The in situ measurements were performed during exposure to the analytes at a total pressure of 0.05–1.05 mbar and 450 K and were correlated with chemiresistor response measurements carried out at a standard pressure and under an ambient atmosphere. The study revealed that heterojunction treatment with ethanol vapors, accompanied by partial reduction of the nanowires, is the key step to obtaining chemiresistors with good sensing performance. While the untreated heterojunctions exhibited poor n-type sensing responses, the treated ones showed significantly improved p-type responses. The treated sensors were characterized by a stable baseline, high reversibility, detection limits estimated as 50 ppm for ethanol and 100 ppb for nitrogen dioxide, and with response times in tens of seconds. In all cases, we propose a band scheme of Cu2O/CuO heterojunctions and a gas-sensing mechanism.
     



    P. Hozák, M. Vorokhta, I. Khalakhan, K. Jarkovská, K. Jarkovská
    J. Cibulková, P. Fitl, J. Vlček, J. Fara, D. Tomeček, M. Novotný, M. Vorokhta, J. Lančok, I. Matolínová, and M. Vrňata
    J. Phys. Chem. C 2019, 123, 49, 29739–29749
    Read more
  4. (2019) Surface Plasmon Enabling Nitrogen Fixation in Pure Water through a Dissociative Mechanism under Mild Conditions

    Nitrogen fixation in a simulated natural environment (i.e., near ambient pressure, room temperature, pure water and incident light) would provide a desirable approach to future nitrogen conversion. As N≡N triple bond has a thermodynamically high cleavage energy, nitrogen reduction under such mild conditions typically undergoes associative alternating or distal pathways rather than follows a dissociative mechanism. Here we report that surface plasmon can supply sufficient energy to activate N2 through a dissociative mechanism in the presence of water and incident light, as evidenced by in-situ synchrotron radiation-based infrared spectroscopy and near ambient pressure X-ray photoelectron spectroscopy. Theoretical simulation indicates that the electric field enhanced by surface plasmon, together with plasmonic hot electrons and interfacial hybridization, may play a critical role in N≡N dissociation. Specifically, AuRu core-antenna nanostructures with broaden light adsorption cross section and active sites achieve an ammonia production rate of 101.4 μmol·g-1·h-1 without any sacrificial agent at room temperature and 2-atm pressure. This work highlights the significance of
    surface plasmon to activation of inert molecules, serving as a promising platform for developing novel catalytic systems.



    C. Hu, X. Chen, J. Jin, Y. Han, S. Chen, H. Ju, J. Cai, Y. Qiu, C. Gao, C. Wang, Z. Qi, R. Long, L. Song, Z. Liu, Y. Xiong
    J. Am. Chem. Soc., Just Accepted Manuscript • Publication Date (Web): 30 Apr 2019
    Read more
  5. (2018) Electrifying model catalysts for understanding electrocatalytic reactions in liquid electrolytes

    Electrocatalysis is at the heart of our future transition to a renewable energy system. Most energy storage and conversion technologies for renewables rely on electrocatalytic processes and, with increasing availability of cheap electrical energy from renewables, chemical production will witness electrification in the near future1,2,3. However, our fundamental understanding of electrocatalysis lags behind the field of classical heterogeneous catalysis that has been the dominating chemical technology for a long time. Here, we describe a new strategy to advance fundamental studies on electrocatalytic materials. We propose to ‘electrify’ complex oxide-based model catalysts made by surface science methods to explore electrocatalytic reactions in liquid electrolytes. We demonstrate the feasibility of this concept by transferring an atomically defined platinum/cobalt oxide model catalyst into the electrochemical environment while preserving its atomic surface structure. Using this approach, we explore particle size effects and identify hitherto unknown metal–support interactions that stabilize oxidized platinum at the nanoparticle interface. The metal–support interactions open a new synergistic reaction pathway that involves both metallic and oxidized platinum. Our results illustrate the potential of the concept, which makes available a systematic approach to build atomically defined model electrodes for fundamental electrocatalytic studies.



    F. Faisal, C.Stumm, M. Bertram, F. Waidhas, Y. Lykhach, S.Cherevko, F. Xiang, M. Ammon,
    M. Vorokhta, B. Šmíd, T. Skála, N. Tsud, A. Neitzel, K. Beranová, K. C. Prince, S. Geiger,
    O. Kasian, T. Wähler, R. Schuster, M. A. Schneider, V. Matolín, K. J. J.
    Nature Materials volume 17, pages 592–598 (2018)
    Read more
  6. (2018) Investigation of gas sensing mechanism of SnO2 based chemiresistor using near ambient pressure XPS

    In this article, we present the results of an investigation into chemical processes which take place at the surface of SnO2-based chemiresistor in various atmospheres (1 mbar of argon, 1 mbar of oxygen, 0.1 mbar of ethanol, 1 mbar of oxygen + 0.1 mbar of ethanol mixture) at common working temperatures (450 and 573 K). The key method for nanoscale analysis was the Near Ambient Pressure X-ray Photoelectron Spectroscopy. In parallel the resistance and DC-responses of SnO2 layer were in-situ monitored providing information about macroscale processes during gas sensing. The change in the sensor resistance after exposure to the ethanol-containing atmospheres together with the disappearance of the band bending effect and observation of different carbonaceous groups including ethoxy groups and acetaldehyde molecules on the sensor surface in the XPS spectra supported the theory of chemical interaction of ethanol with the chemisorbed oxygen. The NAP-XPS spectra also showed that the nanostructured tin oxide is partially reduced even after being exposed to pure oxygen at 573 K. Exposing this surface to the mixture of O2/EtOH did not significantly increase the surface reduction probably due to slow kinetics of the ethanol reduction process and fast kinetics of the oxygen re-oxidation process. However, it was demonstrated that the surface of sensor is slowly getting contaminated by carbon.



    M. Vorokhta, I. Khalakhan, M. Vondráček, D. Tomeček, M. Vorokhta, E. Marešová, J. Nováková, J. Vlček, P. Fitl, M. Novotný, P. Hozák, J. Lančok, M. Vrňata, I. Matolínová, and V. Matolín
    Surface Science, Volume 677, November 2018, Pages 284-290
    Read more

SPARE PARTS

8
Product image
Product description
Article No.
 
Channelplates Set - 1 pair MCP 40

Spare channelplate set for all 40 mm CCD detectors

2060009045
Phosphor screen

Spare phosphor screen for all 40 mm CCD detectors. Exchange with channelplates is recommended.

2060008804
Cu gasket for PHOIBOS 150

PHOIBOS 150 analyzer main flange gasket for Releases R5, R6, R7

2074050088
Nozzle 0,3 mm, double coated

Nozzle 0,3 mm for 7 kV PHOIBOS 150 Backfilling Pre-Lens

2055020988
Nozzle 1,0 mm, double coated

Nozzle 1,0 mm for 7 kV PHOIBOS 150 Backfilling Pre-Lens

2055022579
Rotary feedthrough for IRIS

Replacement feedthrough for PHOIBOS Release R5 & R6 iris mechanism

2060001175
Spindle with Spur and Bevel Gear for Iris

Replacement spindle for PHOIBOS Release R5 & R6 iris mechanism

2055021656
Tubus 3 with Iris

Replacement iris mechanism for PHOIBOS Release R5 & R6

2079150291

DOWNLOADS

6
Close
Your web browser is deprecated
This could effect the presentation and some functions of our website.